Electrical contacts
The electrical contacts, indicated with arrows in the image below, should be carefully cleaned with an alcohol-based cleaner (e.g. IPA) and a
lint-free cloth.
The frequency of cleaning depends on the local environment conditions, but should be at least every 500 probe or stylus holder changes.
RSP3 installation and user's guide
www.renishaw.com
Issued 06 2022