Calibration with universal calibration tower
The universal calibration tower enables calibration of RSP2 and RSP3 probes on a common artefact. The tower is compatible with the REVO
Ø45 mm datum ball for RSP2 calibration, and up to four standard Ø25 mm datum balls for RSP3 calibration. Twelve separate mounting
positions are available, three at each compass point, that allow datum balls to be attached at 45°, 0° or ‐45°.
Periodic requalification should be performed under the following circumstances:
CMM supplier's recommendations, particularly in respect to temperature changes
At the start of the working day or a shift
After accidental collision
When new probe / stylus configurations are introduced
If the initial state is unknown or uncertain
RSP3 installation and user's guide
www.renishaw.com
Issued 06 2022
29