- 18 -
EN
14. Troubleshooting
Problem
Possible cause
Corrective action
ON/OFF switch fails to
work.
•
Power not connected.
•
Device fuse blown.
•
Damaged power cord.
•
Check the power supply.
•
Replace the fuse (refer to Sec. 6.4).
•
Have the device repaired.
Motor fails to start.
•
Motor fuse blown.
•
Motor fault.
•
Replace the fuse (refer to Sec. 6.4).
•
Have the device repaired.
No, or insufficient
vacuum (not in the
green range on the
scale).
•
Intake filter plugged.
•
Seal surfaces dirty.
•
Compressed air hose not
connected, leaky or bent.
•
Venturi nozzle plugged.
•
Intake filter plugged
•
Compressor switched off.
•
Solenoid valve fault.
•
Silencer plugged.
•
Vacuum adjusting knob turned out
too much.
•
Compressed air system with
insufficient performance.
•
Working pressure too low.
•
Air hose too long.
•
Square are of air hose too small.
•
Clean the filter (refer to Sec. 6.1).
•
Clean the seal surfaces (refer to Sec. 6.3).
•
Check the pipe.
•
Have the device repaired.
•
Replace intake filter (refer to Sec. 6.2).
•
Switch the compressor on.
•
Have the device repaired.
•
Replace the silencer.
•
Adjust vacuum (see pt. 5.7).
•
Control the compressor’s performance (see technical
specifications, eventually switch other users off).
•
Increase dynamic working pressure (4,5 to 7 bar /
65,27 to 101,53 psi).).
•
Maximum length 2 m.
•
Minimum inner diameter 4 mm.
Vacuum release /
venting, too slow.
•
Intake filter plugged.
•
Silencer plugged.
•
Solenoid valve fault
•
Clean the filter (refer to Sec. 6.1).
•
Clean risp. replace silencer.
•
Connect a filter unit.
•
Have the device repaired.
Permanent venting
noise.
•
Bowl detection pin sticking.
•
Have the device repaired.
Paddle does not rotate.
•
Paddle is overloaded, motor fuse
(8A) is blown.
•
Clean the paddle and the bowl.
•
Replace the fuse (refer to Sec. 6.4).
Fluttering noise from the
silencer.
•
Working pressure too low
•
Increase dynamic working pressure
(4,5 to 7 bar / 65,27 to 101,53 psi).
We reserve the right to make technical changes.
Summary of Contents for 1824-0000
Page 3: ...0 3546 0 394 inch 1 3 2 4 5 6 7 8 9 10 A E D C B F 11 L M N J H G K 12...
Page 4: ...14 15 16 17 18 19 a b 20 21 a b 22 23 24 a 13 b...
Page 5: ...a b 25 26 27 28...
Page 6: ......
Page 38: ...38 ES...
Page 45: ...45 RU 14 EIN AUS 6 4 6 4 Venturi 6 1 6 3 6 2 5 7 4 5 7 2 4 6 1 8A 6 4 4 5 7...
Page 46: ...46 RU...
Page 52: ...52 ZH...
Page 58: ...58 JA 14 6 4 6 4 Venturi 6 1 6 3 6 2 5 7 4 5 7 2 4mm 6 1 A 6 4 4 5 7...
Page 64: ...64 KO 14 6 4 6 4 6 1 6 3 6 2 5 7 4 5 7 bar 2 m 4 mm 6 1 8A 6 4 4 5 7 bar...
Page 69: ......
Page 70: ......
Page 71: ......