Maintenance
MI3
75
Symptom
J
K
R
S
Temperature over range
> 1200°C
(2192°F)
> 1372°C
(2502°F)
> 1768°C
(3214°F)
> 1768°C
(3214°F)
Temperature under range
-210°C
(-346°F)
-210°C
(-346°F)
-50°C
(-58°F)
-50°C
(-58°F)
Head ambient temperature out of range
> 1200°C
(2192°F)
> 1372°C
(2502°F)
> 1768°C
(3214°F)
> 1768°C
(3214°F)
Table 9: Error Codes for Thermocouple Output TC
Output
Error Code Description
T
−−−
Communication error between head and box
T>>>
Temperature over range
T<<<
Temperature under range
Table 10: Error Codes via Field Bus
Display
Error Code Description
“No sensor”
No sensing head detected
“Sensing head #n lost” Communication error between head and box
“>”
Temperature over top range* e.g. “>600°C”
“<”
Temperature under bottom range* e.g. “<−40°C”
* related to full measurement range
Table 11: Error Codes for LCD Display
11.3
Cleaning the Lens
Keep the lens clean at all times. Care should be taken when cleaning the lens. To clean the window, do
the following:
1.
Lightly blow off loose particles with “canned” air (used for cleaning computer equipment) or
a small squeeze bellows (used for cleaning camera lenses).
2.
Gently brush off any remaining particles with a soft camel hair brush or a soft lens tissue
(available from camera supply stores).
3.
Clean remaining “dirt” using a cotton swab or soft lens tissue dampened in distilled water.
Do not scratch the surface.
For finger prints or other grease, use any of the following:
•
Denatured alcohol
•
Ethanol
•
Kodak lens cleaner
Apply one of the above to the lens. Wipe gently with a soft, clean cloth until you see colors on the
surface, then allow to air dry. Do not wipe the surface dry, this may scratch the surface.
If silicones (used in hand creams) get on the window, gently wipe the surface with Hexane. Allow to
air dry.
1.800.868.7495