Accessories
62
MI3
Figure 49: Air Purge Jacket (XXXMIACAJ)
Figure 50: Mounting the Air Purge Jacket
1.
Remove the sensor
①
and cable from the communication box by disconnecting the wires from
the terminal.
2.
Open the Air Purge Jacket
③
④
and screw the white plastic fitting
②
onto the sensor up to
the end of the threads. Do not over-tighten!
3.
Slip the cable
⑥
through the backside
④
of the jacket.
4.
Close the Air Purge Jacket
③
④
,
reconnect the wires to the communication box and apply the
mounting nut
⑤
.
10.2.4
Air Cooling System
The Air Cooling System can not be combined with LTH heads!
The sensing head can operate in ambient temperatures up to 200°C (392°F) with the air-cooling
system. The air-cooling system comes with a T-adapter including 0.8 m / 31.5 in (optional: 2.8 m / 110
in) air hose and insulation. The T-adapter allows the air-cooling hose to be installed without
interrupting the connections to the box.
The air-cooling jacket may be combined with the right angle mirror.
Hose with inner
diameter of 3 mm
(0.12 in), outside
5 mm (0.2 in)
Fitting to M5
inner thread