Operation
48
MI3
<Laser>
handles the laser in the following modes:
<off> switches the laser off
<flash> forces the laser to blink at 8 Hz
<on> switches the laser permanently on
<external> switches the laser via external input FTC3
An activated laser will be switched off automatically after 10 minutes.
The laser is available for 1M and 2M heads only. The laser can be activated at
the same time for maximal 4 heads.
<Average> AVG
signal post processing set to averaging, parameter given in seconds. Once
<Average> is set above 0 s, it automatically activates. Note that other hold
functions (like Peak Hold or Valley Hold) cannot be used concurrently. Value
range: 0.0 to 998.9 sec,
∞
See section 8.5.1
<Peak Hold> PH
signal post processing set to Peak Hold, parameter given in seconds. Once
<Peak Hold> is set above 0 s, it automatically activates. Note that other hold
functions (like Valley Hold or Averaging) cannot be used concurrently. Value
range: 0.0 to 998.9 sec,
∞
See section 8.5.2
<Val. Hold> VH
signal post processing set to Valley Hold, parameter given in seconds. Once
<Valley Hold> is set above 0 s, it automatically activates. Note that other hold
functions (like Peak Hold or Averaging) cannot be used concurrently. Value
range: 0.0 to 998.9 sec,
∞
See section 8.5.3
<Trigger>
defines the trigger mode for the selected head:
<trig>: … to reset the peak or valley hold function
<hold>: activates the hold function
See section 7.4
<Alarm Mode>
defines the alarm mode for the selected head:
<Tobj>: object temperature as alarm source
<Tamb>: head ambient temperature as alarm source
<Set Point>
defines a temperature threshold for an alarm
<Lo Limit>
low end of temperature measurement range (read only)
<Hi Limit>
high end of temperature measurement range (read only)
<Fact. default>
sets the selected head back to factory default.
The factory default values are to be found in section 18.9
page 108.
<Type>
provides the head model, e.g. MI3LT
<SN>
provides the serial number for the selected head and allows to reassign a new
head address
8.3
<Box Setup> Page
<Relay Mode>
defines the switching behavior for the box internal alarm relay:
<normally open>: open contact in non-alarm status
<normally closed>: closed contact in non-alarm status
<permanently OFF>: permanently open contacts
<permanently ON>: permanently closed contacts