Maintenance
76
MI3
Do not use any ammonia or any cleaners containing ammonia to clean the lens. This
may result in permanent damage to the lens’ surface!
11.4
Sensing Head Exchange
To exchange a sensing head, the following procedure is required:
1.
Disconnect power to the box.
2.
Disconnect all head wires from the box terminal.
3.
Power the box.
4.
The alarm indicator of the box starts to blink indicating a lost sensing head.
5.
Press the
button to navigate to the head page indicating a lost sensing head.
6.
Decide whether you want to select:
a) <Remove Yes>: to
remove
the head permanently from the box by loosing all head
parameters (head address de-allocated for other heads, alarm condition is reset) – the next
head connected later will be detected as a new head and automatically assigned to a free head
address.
or
b) <Remove No>: to
keep
the head assigned to the box by saving all head parameters for a
future use without the need to parameterize that head again (head address reserved for that
individual head, alarm condition is kept) – the same head connected later will be detected as a
known head by keeping the previous address.
1.800.868.7495