After
all
the
necessary
snapshots
are
taken,
wait
for
the
result
of
calibration
procedure.
Resulting
accuracy
of
calibration
is
specified
in
pixels
.
A
good
result
is
accuracy
no
worse
(not
higher)
than
0.05
pix
.
If
you
get
a
higher
result,
repeat
the
calibration,
carefully
checking
the
plates
position
and
brightness.
If
is
is
necessary,
repeat
calibration
without
blue
light
and
filters.
While
working
with
the
scanner,
don't
forget
to
once
in
a
while
perform
orientation
-
it
also
serves
as
an
accuracy
indicator.
If
the
accuracy
during
orienting
starts
to
be
very
different
from
the
original
one,
you
need
to
re-calibrate
the
equipment.
Some
advice
on
calibration:
1.
Do
not
change
the
distance
from
the
center
of
the
plate
to
the
scanner
when
turning
the
plate.
The
exception
are
only
snapshots
in
positions
10
and
11.
2.
Do
not
rotate
the
plate
at
a
very
high
angle.
Ensure
that
all
marks
on
the
plate
are
visible
from
both
cameras,
when
you
turn
the
table.
3.
Be
careful
with
calibration
plates!
It
is
not
permitted
to
contaminate
or
perform
any
mechanical
damage
to
the
surface
with
the
marks.
After
the
use
store
the
plates
in
designated
case.
4.
Before
the
calibration
procedure
ensure
that
cameras
and
cables
are
secure
in
the
designated
position.
Fast calibration