2.
In
accordance
with
the
text
prompt
and
symbol
image
install
the
calibration
plate
in
the
desired
position.
The
brightness
of
the
images
of
calibration
plate
can
be
adjusted
with
exposure
controller.
You
should
not
allow
very
dark
images
or
images
with
excessive
brightness
areas.
Press
the
Capture
button.
Original
plate
position
-
the
position
at
the
working
distance
from
the
scanner
(marking
on
the
view
from
cameras
matches
the
markers
on
the
plate),
the
cross
is
projected
on
the
central
marker.
3.
If
the
plate
is
placed
incorrectly,
a
message
will
appear.
Correct
the
placement
and
repeat
the
capture.
In case if not all marks are found on the image, you will also see an error message. It
may mean the following:
●
not all markers are visible on one of the snapshots,
●
not all markers are lit by the projector,
●
calibration plate is mounted at large angle to the scanner,
●
the plate is too close or too far from the scanner - the image is not sharp,
●
the plate is damaged or dirty.
Please
correct
all
deficiencies
and
press
Capture
.