Adjust the lenses diaphragms. Set the exposure for the cameras between 30 and 60
(less if the object, which you are planning to scan is dark and more if it is bright). By
default the value is 45.
Areas
illuminated
with
red
mean
overexposed
areas
with
too
bright
image.
Configure lenses iris so that the lines on the calibration plate are as bright as possible,
but without a significant amount of excessive light. Image brightness of both cameras
must be the same!
The lens focus is set during the scanner manufacturing and does not need to be
adjusted by the user.