Select scanning zone
After
preparatory
actions
on
setting
up
the
computer,
you
should
carry
the
scanner
setting
procedure.
Select
Configure
scanner
for
a
new
area
in
Calibration
window
on
the
Start
screen
and
follow
the
instructions
for
each
step.
Before beginning the configuration, select the desired scan area (zone), depending on
the object scanned. Scanning area is a set of interchangeable optics, calibration plate
and stand for the calibration plate, which help to configure the scanner to scan a certain
area. In the majority of cases, scan area is correctly matched, if the size of the object
approximately equals to the size calibration plate.
Calibration field parameters
Take the necessary calibration to configure the scanner to the selected scan area.
Make sure that the scan area is included in the package of your scanner.
The exact size and measurement temperature are
written on the label on the reverse side of the calibration
plate.
Enter these values into the
Configuration Wizard
or
select from the drop-down list of previously entered
values.