PROGRAMMING OF MULTIRAE
4 - 44
4.8.2 Enable / Disable Sensor
This function allows the user to selectively enable or disable
individual sensors in the
MultiRAE PLUS
monitor. When
a sensor is disabled, the sensor will not measure nor display
the gas concentration.
1. “Enable / Disable Sensors?” is the second sub-menu item
in Table 4.7.
2. Press
[Y/+]
key. The display shows all the installed
sensors in the monitor. A previously enabled sensor is
marked by an “*’ next to the sensor name. The cursor
will blink at the first sensor location. Press
[Y/+]
key to
enable the sensor and
[N/-]
key to disable the sensor. A
previously disabled sensor will show “----” instead of the
sensor name.
TOX1
VOC
TOX2
CO* VOC
H
2
S*
LEL* pick OXY*
LEL
OXY
3. Press
[MODE]
key momentarily to move from one
sensor location to the next one. Repeat Step 2 until all
the sensors which needs to be enabled are selected.
Press and hold
[MODE]
key for 1 second to save the
new sensor selection.
4. The display shows “Save ?” To confirm the new selection,
press the
[Y/+]
key to accept the change. Press the
[N/-]
key or the
[MODE]
key to discard the change and move
to the next sub-menu.