COMPUTER INTERFACE FOR MULTIRAE PLUS
5 - 17
This is the place where the user can specify the low alarm
level (%) and high alarm leve l (%) for OXY sensor.
VOC Alarm Levels (ppm):
This is the place where the user can specify the STEL alarm
level (ppm), TWA alarm level (ppm), low alarm level (ppm)
and high alarm level (ppm) for VOC sensor.
Tox1 Alarm Levels (ppm):
This is the place where the user can specify the STEL alarm
level (ppm), TWA alarm level (ppm), low alarm level (ppm)
and high alarm level (ppm) for Tox1 sensor.
Tox2 Alarm Levels (ppm):
This is the place where the user can specify the STEL alarm
level (ppm), TWA alarm level (ppm), low alarm level (ppm)
and high alarm level (ppm) for Tox2 sensor.
Alarm mode:
If
Auto
reset is chosen, the alarm will stop as soon as the
alarm condition goes away. The alarm will remain on if the
Latched
option is chosen. To turn off a latched alarm, press
[Y/+]
key to acknowledge the alarm condition.
Power On Zero:
This option allows the monitor to perform a fresh air
calibration automatically when the monitor is turned on.
Care must be taken, when using this mode, to ensure that the
monitor is in a fresh air environment when turned on.
Calibration span: