MiniRAE Lite User’s Guide
31 Maintenance
The major maintenance items of the instrument are:
•
Battery pack
•
Sensor module
•
PID lamp
•
Sampling pump
•
Inlet connectors and filters
Note:
Maintenance should be performed by qualified personnel
only.
NOTE:
The printed circuit board of the instrument is connected to
the battery pack even if the power is turned off. Therefore, it is very
important to disconnect the battery pack before servicing or replacing
any components inside the instrument. Severe damage to the printed
circuit board or battery may occur if the battery pack is not
disconnected before servicing the unit.
Battery Charging & Replacement
When the display shows a flashing empty battery icon, the battery
requires recharging. It is recommended to recharge the instrument
upon returning from fieldwork. A fully charged battery runs a
instrument for 16 hours continuously. The charging time is less than 8
hours for a fully discharged battery. The battery may be replaced in
the field (in areas known to be non-hazardous), if required.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in area known to be non-hazardous
.
Remove and
replace battery only in areas known to be non-hazardous.
52