MiniRAE Lite User’s Guide
Measurement range & resolution
Lamp Range Resolution
10.6 eV
0.1 ppm to 5,000 ppm
0.1 ppm
Response time (T
90
):
2
seconds
Accuracy
10 to 2000 ppm:
±
5% at calibration point.
(Isobutylene):
PID Detector:
Easy access to lamp and sensor for cleaning
and replacement
Calibration:
Two-point field calibration of zero and
standard reference gases
Inlet Probe:
Flexible 5" tubing
Keypad:
1 operation key and 2 programming keys; 1
flashlight switch
Direct Readout:
Instantaneous, peak value, and battery
voltage
Intrinsic Safety:
Europe: ATEX (II 1G EEx ia IIC T4)
IECEx (Ex ia IIC T4)
EM Interference:
Highly resistant to EMI/RFI.
Alarm Setting:
Separate alarm limit settings for Low Alarm
and High Alarm
Alarm:
Buzzer 95dB at 30cm and flashing red LEDs
to indicate exceeded preset limits, low
battery voltage, or sensor failure
Alarm Type:
Latching or automatic reset
Real-time Clock:
Automatic date and time
Communication:
Download instrument setup from PC via RS-
232 with Travel Charger, or via USB with
optional Charger Stand
Sampling Pump:
Internally integrated. Flow rate: 450 to 550
cc/min.
10