- 87 -
18.5.
Means of carrying out drying process
Bottom section of moisture analyzer’s main display features three pictograms,
which indicate current drying mode settings.
If these parameters are incorrect, they should be changed in a way , that are
correct for a dried sample that is tested.
Means of setting drying parameters is described in the previous point of this user
manual.
If drying process is based on a drying program from the database, then:
•
Enter submenu < Databases>, and press <
Database of drying
programs>
•
Select drying program by pressing field with its name. The moisture
analyzer returns to displaying the main window. The grey coloured
workspace indicates name of selected drying program, and pictograms on
drying parameters indicate data of enabled drying program.
•
On enabling a drying program, the drying settings cannot be changed.
Drying parameters can only be changed if no drying program is enabled.
•
An additional option is displaying (in an enabled drying program)
descriptions on additional equipment needed for correct carrying out the
drying process, sample preparation and recommended sample size. Using
the options requires pressing the <Sample description> button in the
bottom bar which opens a window with descriptions entered for the drying
program.
Caution:
The option is available only on selecting an optional drying program from the
<Database of drying programs >.