- 72 -
-
User defined 2 (set moisture content change ratio
∆
%M over time 60s)
-
Test (enables adjusting finish mode parameters to a tested sample)
Automatic finish mode – user defined criterion (mass/time)
MA 3Y operator determines the limit value for loss of mass and time in which
the loss of mass must not exceed a pre-set value. As the moisture analyzer
reaches the set criterion it automatically finishes the measurement (drying
process).
Automatic finish mode – user defined criterion (change of moisture
content)
An operator determines limit loss of moisture content over time of 60s. As the
moisture analyzer reaches the set criterion it automatically finishes the
measurement (drying process).
Manual finish mode
An operator ends drying process manually by pressing STOP soft key on
moisture analyzer’s display.
Time defined finish mode
Drying process finish takes place on elapsing a set amount of time,
independently on measurement result.
18.
DRYING PROCESS
18.1.
Working mode activating procedure
•
while in the main window, press
soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of
available working modes,
•
select <
Drying
> mode, the software automatically returns to the
main window and displays
pictogram in the upper bar,
The grey coloured workspace contains the following data:
o
Drying mode
o
Product
On selecting the drying mode , the display contains the following
quick access keys in the bottom bar:
Summary of Contents for MA 3Y