- 144 -
Format:
BP_TIME <CR><LF>
Accessible responses:
BP_OK <CR><LF>
– command completed, BEEP signal activated
BP_E” <CR><LF>
– incorrect format or lack of the parameter
BP_I <CR><LF>
– command understood but not accessible at this
moment
TIME
– parameter, decimal value, determined time limit for the signal, given
in [ms]. Recommended range <50
÷
5000>.
If value greater than the valid one is given, the BEEP sound is activated for
the maximal available time limit.
Example
Command:
BP_350<CR><LF> – turn BEEP signal for 350 ms
Response:
BP_OK<CR><LF> – BEEP on
Caution! – BEEP signal activated by BP command gets aborted if it is activated
by means of other devices: keyboard, touch panel, proximity sensors.
OMI - list available working modes
The command lists available working modes for a given device
Format:
OMI <CR><LF>
Accessible responses:
OMI <CR><LF>
n_”Mode name”
<CR><LF>
:
n_”Mode name”
<CR><LF>
OK <CR><LF>
– command completed, available working modes are
given
OMI_I <CR><LF>
– command understood but not accessible at this
moment
Mode name
– parameter, working mode name, given in between inverted
commas. Form of the name is the same like on the balance display, it is given
in a currently operating language.
n
– parameter, decimal value specifying working mode number.
n
→
1 – Weighing
17 – Pippette calibration
Caution: working modes numbering is assigned to working modes names and
it is the same for each balance types.
Some balances in response to
OMI
may give working modes numbers without
names.
Example1
Command:
OMI <CR><LF>
– give available working modes