- 122 -
21.3.3 Test
<
Test>
function enables correct operation of barcode scanner connected to
balance.
Procedure:
5.
Open
< Barcode scanner>
submenu.
6.
Press
<
Test>
parameter;
<Test>
window opens with ASCII and HEX fields
in it.
7.
Scanned barcode is entered into ASCII or HEX field; result of test is displayed in
the bottom part of the window.
When:
•
<Prefix>
and
<Suffix>
are identical for both – balance settings and scanned
barcode – the test is
<Positive>.
•
<Prefix>
and
<Suffix>
differ for both – balance settings and scanned barcode –
the test is
<Negative>
.
21.4.
Transponder card scanner
Selecting (logging) a user on each switching on the moisture analyzer can be
carried out by means of:
•
Inserting user’s password using on-screen keyboard,
•
With a transponder card scanner, by approximating a registered card to the
scanner.
Caution:
For correct moisture analyzer’s cooperation with a transponder card scanner set
appropriate baud rate value in submenu <
Communication> (default 9600b/s).
Transponder card scanner port.
The moisture analyzer enables communicating with a transponder card
scanner using the following ports:
•
RS 232 (COM1),
•
RS 232 (COM2).
Assigning number of a card to a user:
In order to log in an operator using a transponder card scanner, first assign a
number of an already registered card to a selected user in the database of
operators.
Procedure:
•
Connect a transponder card scanner to a moisture analyzer’s port (RS 232
COM1 or RS 232 COM2), select communication port in moisture analyzer
settings for the transponder card scanner
•
In submenu <
Communication> set baud rate for compatible with the
one of the transponder card scanner (default 9600b/s).