- 46 -
After 8 minutes, the system
opens a window with an on-
screen numeric keyboard.
Using the on-screen keyboard
insert the current temperature
value of the moisture analyzer,
that is indicated on the
external control thermometer.
Press <Accept> key to proceed
to the following adjustment
step.
The moisture analyzer
proceeds to the last step of
adjustment process. It heats
the drying chamber to a higher
temperature and for another 8
minutes it maintains this
temperature and displays its
value.
After 8 minutes, the system
opens a window with an on-
screen numeric keyboard.
Using the on-screen keyboard
insert the current temperature
value of the moisture analyzer,
that is indicated on the
external control thermometer.
Press <Accept> key to proceed
to the following adjustment
step to complete temperature
adjustment procedure.
On completing temperature adjustment procedure, the moisture analyzer
displays a corresponding message box. Accept the message box by pressing
<ACCEPT> key and return to the main window.