- 148 -
carrying out, no parameter
available or incorrect format
EV_I <CR><LF>
- command understood but not
accessible at this moment
n
– parameter, decimal value determining ambient conditions state
n
→
0 – stable ambient conditions
1 – unstable ambient conditions
Caution:
Command changes settings for a current working mode.
Example:
Command:
EV_1 <CR><LF> – set value ‘stable’ for ambient conditions option
Response:
EV_OK <CR><LF> – ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until
command EV 0 swaps it to value <UNSTABLE>.
FIS – Set filter
FIS – (Filtetr, Set)
Command overview:
Command lets the user set filter for a particular device.
Format:
FIS_n <CR><LF>
Accessible responses:
FIS_OK <CR><LF>
- command carried out
FIS_E <CR><LF>
- an error occurred on command
carrying out, no parameter
available or incorrect format
FIS_I <CR><LF>
- command understood but not
accessible at this moment
n
– parameter, decimal value determining filter number
n
→
1 – very fast
2 – fast
3 – average
4 – slow
5 – very slow
Caution:
The numbering is assigned to a particular filter name and it is identical for all
balance types.
The command changes settings for a current working mode if, for a particular
balance type, filter settings are assigned to the working mode.