- 142 -
ES CR LF
- command not recognized (incorrect mass format)
Set high checkweighing threshold HI
Format:
UH_XXXXX CR LF
, where: _ - space,
XXXXX
– mass format
Accessible responses:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognized (incorrect mass format)
Give value of low checkweighing threshold HI
Format:
ODH CR LF
Accessible response:
DH_MASS CR LF
– command carried out
Response format:
1 2
3
4-12
13
14 15 16
17
18 19
D H
space
mass
space
unit
space
CR LF
Mass
- 9 characters with right justification
Unit
- 3 characters with right justification
Give value of high checkweighing threshold LO
Format:
OUH CR LF
Accessible response:
UH_MASS CR LF
- command carried out
Response format:
1 2
3
4-12
13
14 15 16
17
18
19
U H
space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
User log in
Format:
LOGIN Name, Password CR LF
(Name and password have to be
inserted as they are set in the moisture analyzer – small and capital letters)
Accessible responses:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERROR CR LF
- command understood, an error in name or
password occurred, log in failed
ES CR LF
- command not recognized (error in format)