9
2.6. Staff training
This balance may only be operated and looked after by trained member of staff.
3. TRANSPORT AND STORAGE
3.1. Acceptance check
Please check the packaging immediately upon delivery and the device during
unpacking for any visible signs of external damage.
3.2. Packaging
Please retain all parts of the original packaging in case it should be necessary to
transport or return items at any time. Only the original packaging should be used
for return consignments. Before dispatch, disconnect all attached cables and
loose/movable parts (weighing pan, protecting shields, etc). Please place any
elements of the balance in its original packing. Please protect the balance and its
parts against any damage while in transport.
4. UNPACKING, INSTALLATION AND COMMISSONING
4.1. Installation and operation place
* Balance should be stored and used in locations free of vibrations and
shakes, free of wind and dust.
* Air temperature should not exceed range between:
+10
o
C
÷
+45
o
C
* Humidity of air should not exceed 80% (not condensing).
* When operating the balance, room temperature should not change more
than 0,5°C within one hour,
* The balance should be placed on a stable surface not affected by vibrations
and distant from heat sources.
* Please take special safety measures when weighing magnetic loads, as part
of this balance is a strong magnet. If there is a necessity to perform
measurement of such load, please use option of under hook weighing, which
removes the load from area influenced by the magnet. For installation place
of under hook device, please look at the bottom casing of the balance,.
*
Avoid static charging of items to be weighed, or weighing container. In case
of static discharges may occur, please ground the device. Grounding bolt is
placed At the back side of the balance’s casing.