73
Function
Format
S CR LF
(sending results now)
Function CONSTANCE
WORK
- equivalent for key conA on
Format
C 1 CR LF
(start continuous transmission off in basic interval)
Function MANUAL
WORK
- equivalent for order conA OFF
Format
C 0 CR LF
(To change work mode on manual printout)
Function CONSTANCE
WORK
- equivalent for key conb on
Format
CU 1 CR LF
(start continuous transmission off in actual interval)
Function MANUAL
WORK
- equivalent for order conb OFF
Format
CU 0 CR LF
(To change work mode on manual printout)
Function
SEND THE RESULT IN ACTUAL INTERVAL
Command
SU CR LF
(result in actual unit is sent from the balance after
stability)
Function
SEND RESULT IN ACTUAL INTERVAL IMMEDIATELY
(result in
actual unit sending results now)
Command
SUI CR LF
If command which is not included in the register or is an error and ends with CR LF
is sent to the balance the balance returns command E S CR LF.
18. ERROR MESSAGES
Er1 Hi
- wrong initial mass or calibration
Er2 nuLL
- values from A.D converter out of the normal range - under range
Er3 FuL1
- values from A.D converter out of the normal range - over range
Er4 FuL2
- over range
Er5 rout
- value out of range for performed function e.g value > +/- 2% from initial
mass for zeroing,
e.g. value <= 0 for tarring
Er7 tout
- timeout error for zeroing and tarring
Er8 outr
- enter value out of range
Er9 Lock
- function locked e.g. to protect function if LFt (legal for trade) is on,
Er10 cal
- calibration error (change of mass or incorrect mass of standard mass).