Available options:
•
METHOD
It is a means for designating an accepted quality standard.
Pressing a field displays a window with a list of minimal sample
weight values stored in balance’s memory, following a criterion on
determining the minimum sample.
Adding a new method is available only from the menu level
<Databases /Minimum samples>
•
MODE
Lock
– on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. Additionally the software disables accepting the measurement
result if it is below the minimum sample limit
Warn –
on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. The user can accept a measurement that is below the
minimum sample limit, but on a printout such measurement is
preceded by an asterisk (*).
CAUTION:
Means of setting a method of minimum sample determining is described
in point 27.9. of this user manual: <Databases /Minimum sample>.
RADWAG authorized personnel is only authorized to add new values of
minimum samples.
Weighing with application of <MINIMUM SAMPLE> mode
If during a weighing process a user wants to be informed whether a
measurement exceeds the minimum sample limit for a specific weighing
threshold, then the <Minimum sample> mode has to be enabled in the
additional setting of the weighing mode.
Procedure:
1.
Press grey coloured workspace
2.
The display indicates menu: Settings, Keys, Information, Printouts and
Profiles
3.
Select menu <Settings> - additional options on weighing process
4.
Press < Minimum sample> field
5.
Below window is opened