- 170 -
29.11.
Reports from pipette calibration processes
Reports contain data on completed pipette calibration processes. In case of
multichannel pipettes, the reports are generated for each channel separately.
Each report can be previewed, searched by date, exported or printed.
Procedure
•
Enter submenu < Databases>, and press <
Reports from pipette
calibration processes >
•
Press field of a desired report record, unless it is visible on the top of the
list, wind the list forward using navigating keys
•
Report’s name consists of date and time of its creation,
e.g.: 2012.03.12 11:12:15
Caution:
It is possible to use an option for searching a report.
List of parameters in a report on pipette calibration process
:
1.
Pipette [name from the database of pipettes]
2.
Serial no. [serial no. inserted during calibration process]
3.
Channel no.
4.
Start date
5.
End date
6.
User [name of a user carrying out calibration]
7.
Client [client name]
8.
Number of measurements
[no. of measurements for each tested
volume]
9.
Operation with ISO 8655
[information if errors comply with the
standard]
10.
Status [information whether the errors are below permissible
value accepted for the tested volume]
11.
Temperature [average temperature value during calibration
process]
12.
Humidity [average humidity value during calibration process]
13.
Pressure [average pressure value during calibration process]
14.
Water temperature [average water temperature during
calibration process]
15.
Z coefficient [value if the Z coefficient for given temperature]
29.12.
Series
Database of series contains data saved for a series and measurements of
samples.
Procedure
•
Enter submenu < Databases>, and press <
Series> field
•
Press <
Add > key to add a new series to the database