- 194 -
33.6.
Sensor sensitivity
The parameter is adjustable between 0 – 9, and it determines the
distance from which the IR proximity sensors should react.
As standard, the sensitivity is set within the range of 5 to 7.
Procedure:
•
Enter submenu<
Others>
•
Select parameter<Sensor sensitivity>which opens an editing window,
•
Select one of available options. The selection causes immediate returning
to the menu window.
33.7.
Autotest
AUTOTEST function is designed to aid a user in both assessing balance’s
operation and diagnosing the reasons for occurrence of errors in weighing
which exceed the maximum permissible values for a given balance
model.
By means of a simple, repeatable and fully documented way the function
enables optimizing balance’s settings to maintain the best possible
repeatability and weighing time at a workstation. The main purpose of
function operation is the possibility of monitoring the above mentioned
parameters at optional moment and saving records from carried out
tests.
The functions is divided into two modules:
AUTOTEST FILTER
AUTOTEST GLP
Before carrying out any of the tests , the balance controls its level status,
temperature and humidity.
AUTOTEST FILTER
It is process of 10 times loading and unloading of the internal weight for
all possible settings of balance’s filter and value release values. The test
controls 2 parameters:
•
Repeatability;
•
Stabilization time of measurement result.
The whole testing process takes approximately 1 hour. On completing the
test for all possible settings, balance displays a message box with
obtained results.
The user is given information on optimum balance settings in present