RS-
330/332 Operators’ Manual – Revision 2.00
SURVEY
Radiation Solutions Inc
– Proprietary
Page 14
Revision Date: May 16, 2017
Customers Recording Data in OnTheFly Assay mode may find that if the device had been previously
setup for Record Mode that the sequential order or chronological numbering for saved files will be
affected, as the device is also storing data for OnTheFly storage. It appears that the sequential
numbering is skipping as the device is storing data on the fly. Each Record assay and OnTheFly assay
is assigned a unique ID. From affected customer`s data it appears that assays were being saved while
the “OnTheFly” option was started. In this case the unit would assign an ID to the Manual (Record)
assay and skip the OnTheFly assay, thus resulting in non-sequential ID numbering. To prove this is
happening compare the time stamps of the exported “Manual assay” data and the “OnTheFly assay”
data.
When you export the data the “OnTheFly assays” will be the ones that have a blank space in the
total column while the Manual assays will contain all the data.
D-1036 REV 02.00 Status: RELEASED printed 16/05/2017 9:45:26 AM by Radiation Solutions Inc.