7
Do not use around flammable materials
To avoid possible fire or explosion hazard, do not use around
the following situation/place.
close to flammable materials
close to hyperbaric oxygen tank (within 25m)
close to hydrogen tank (within 25m)
Do not attempt to recharge or cause heat/damage the battery
(BT-303R)
Recharge or heat/damage on battery(BT-303R) may
cause an explosion or fire hazard. If the battery is low, please
exchange to new one
Do not attempt to disassemble or repair the AED
Disassembling AED may result electric shock hazard.
RADIANQBIO is not responsible for electric shock hazard
resulted by unauthorized disassemble or repair. Please
contact local/national sales representative for repair or
replacement
Electromagnetic emissions and equipment adjacent
• Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that
they are operating normally.
• Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
• Portable RF communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part
of the HR-501, including cables specified by the
manufacturer.