26
6.1
Self-diagnosis
Power-On Self-test
(Operation mode)
Every time when power button is pressed
Flash ROM test
Charge/Discharge test
Battery Output level test
Regular Self-test
(Stand-by mode)
Automatically performed once per month on 1
st
day
Flash ROM test
Charge/Discharge test
Battery Output level test
If the product fails in self-diagnosis test, status light blinks in red and beeper sounds
a warning. In this state, the product does not operate properly. If the product does
not operate properly, users should refer to 6.3. Troubleshooting in the manual.
During periodic self-diagnosis, if product trouble is found, status light blinks in red;
in case of power on by pushing power button, beeper sounds 3 times every second
for warning with error audio guidance.
6.2
AED conditional alarm