33
7.8
Guidance and Manufacturer’s Declaration
Guidance and Manufacturer’s Declaration – Electromagnetic emissions
Emissions test
compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
(EN 55011)
Group 1 ,
Class B
The HR-501 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The HR-501 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable Using of battery
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable Using of battery
Guidance and Manufacturer’s Declaration – Electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
EN/IEC 61000-4-2
±6 kV contact
±2, ±4, ±8, ±15 kV air
±6 kV contact
±2, ±4, ±8, ±15
kV air
The HR-501 uses interference
detected and motion detected
indicators to notify the user if
conditions are not ideal. No other
ESD requirements are necessary.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power line supply
lines
±1 kV for input/output lines
100 kHz repetition rate
Not applicable
Using of battery
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not applicable
Using of battery
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
100% reduction for 0.5
cycle at 0, 45, 90, 135,
180, 225, 270 and 315
degrees
100% reduction for 1
cycle
, 30% reduction for
25/30 cycles at single
phase at 0 degrees
100% reduction for
250/300 cycles (5 sec.)
and
all devices remain
safe
Not applicable
Using of battery
Power frequency (50/60
Hz)
magnetic field
EN/IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
30 A/m
50 Hz and 60
Hz
Power frequency magnetic fields
should not be greater than levels
characteristic of a typical location in
a commercial or hospital
environment.
Radiated RF
EN/IEC 61000-4-3
3 V/m, 10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
3 V/m, 10 V/m
80 MHz to 2.7
GHz
80% AM at 1
kHz
Table 9 in
Portable and mobile RF
communications equipment should be
used no closer to any part of the HR-
501, including cables, than necessary.
The recommended separation
distance calculated from the equation
applicable to the frequency of the