Guidance for maintenance
5.1
AED location and periodical maintenance
5.1.1
Please set AED at a place where AED can be easily found and accessed.
RADIANQBIO strongly suggests setting AED close to a telephone in order to
call emergency aid service immediately in case of emergency
5.1.2
All components including P-303 electric pads shall be kept together
Do not sterilize the AED or its accessories.
5.1.3
Always, battery shall be installed to AED
5.1.4
Periodically check AED operation and expiration date of battery and AED pads
5.1.5
The product shall be stored/operated in following environmental conditions
①
Stored/operated in following environmental conditions for AED
Temperature Range : 0℃ to 50℃ (32 to 122℉)
Humidity : 5% to 95% (non-condensing)
②
Stored temperature range of Shipping/transporting (within 5days)**
Temperature Range : -20℃ to 60℃ (-4℉ to 140℉)
Humidity : 5% to 95% (non-condensing)
③
Avoid AED storage in below places
Place where wide temperature changes happen frequently
Place where is exposed to direct sunlight
Wet or flood places
Places close to heater or heavy electrical devices
Places where excessive shock or shake may occur
Places close to flammable materials
Places exposed to the extreme environmental weather condition
5.1.6
Battery and AED pads maintenance
①
Battery maintenance
If the battery is low, AED may not be operated properly in emergency
situation. When status LED blinks ‘
RED
’, battery pack shall be replaced
immediately. Periodically
check manufacturing date indicated on battery
label, and replace to new one if it is expired (4 years from manufacturing
date)