30
7.3
Functionality
Parameter
Detail
FFT ECG Analysis
The AED FFT ECG analysis system automatically
analyzes the patient’s ECG and advises you when
defibrillation is required
Waveform
Biphasic Truncated Exponential (BTE)
Impedance range
25Ω to 175Ω
Electric shock energy
Adult mode: 150J±10%
Child mode: 50J±10%
Charge time
Since the initiation of AED operation (Power On),
8 seconds with a new, fully charged battery; with low-
on battery, the charge time gets significantly longer
Operating time
For a new, fully charged battery pack at 20~ 30°C:
200 discharges at maximum energy or 10 hours of
continuous ECG monitoring.
Automated Self-tests
Monthly (every 1
st
day for each months):
Battery, AED pads, internal electronics
Audible alerts
Voice recorded prompts
Alarm beep sound
Indicators
Overall status light (Battery, AED condition)
CPR procedure guidance LED
CPR procedure guidance draws
ECG data record
13.75Mb (100 of ECG logs can be recorded)
Data communication
Bluetooth 2.0