The general guidelines below are relevant for all SpaTouch Elite applications.
To insure client safety, attain optimal clinical results and protect device
performance, operators should adhere to the following protocols.
Activate –
Turn on the system
Set Energy Levels –
Always use the test energy levels defined in
Table 4 as a starting point.
Position Handpiece–
Always position the Handpiece in full contact
with the skin with 15-20% overlapping. After the pulse is administered,
immediately and completely remove the handpiece from the skin.
Apply Pulse –
Press the footswitch. With the footswitch pressed,
simultaneously push the handpiece pulse switch.
Rest Between Pulses –
Remove the handpiece completely away
from the skin to prevent preheating the skin.
Apply Additional Pulses –
To trigger the next pulse, wait until the
“Ready” indicators are lit and the audible signal is heard. Place the
Handpiece on the next treatment spot and repeat.
Overlapping –
Overlap pulses between 15-20%. Stacking (placing
one pulse on top of the other) and/or excessive overlapping can cause
crusting or blisters which can lead to temporary pigmentary changes.
Area Adaptors –
The Area Adapters should be used when treating
small or narrow areas. No energy level adjustments are necessary
when using the small area adaptors.
Hygienic Rings –
Use a new hygienic ring for each client. You may
reuse the same hygienic ring for the same client. Remember to clean it
after each use.
WARNING
: To avoid eye discomfort, never look directly at the light emitted
from the Handpiece.