If a system error occurs, one of the following appears on the work screen:
E1, E2, E3, E4, E5, E6, E7, E8, E9, F1, F2, F3, F4, F5, F6 or F7.
In all instances, begin by turning off the system using the main power switch
located on the back.
Wait 5 minutes and then turn on the system. If the error persists, follow the
steps mentioned below. Do not remove the cover to open the system and
attempt to fix the problem yourself. This will void your warranty and/or
aggravate the problem.
Table 6: List of System Errors
Error
Fault
Solution
E1
Blower fault
Shut down the system using the power switch on
the back of the system.
Contact Radiancy or a Radiancy authorized
technician.
E2
Fan fault
Make sure nothing is physically blocking the fan.
Shut down the system using the power switch on
the back of the system. Wait 5 minutes, and then
restart the machine.
If nothing is blocking the fan and the problem
persists, shut down the system and contact
Radiancy or a Radiancy authorized technician.
E3
Handpiece is
not properly
connected
Shut down the system using the power switch on
the back of the system and wait 5 minutes.
Detach the handpiece and then reattach it, making
sure it is properly connected.
Turn the system on once again using the main
power switch.
If the problem persists, shut down the system and
contact Radiancy or a Radiancy authorized
technician.
E4
Charging fault
Shut down the system using the power switch on
the back of the system.
Contact Radiancy or a Radiancy authorized
technician.
E5
Unrecognized
handpiece
Fault
Make sure you are using the SpaTouch Elite
handpiece (and not a Mistral or Kona
handpiece)
If the problem persists, follow the steps for
solution indicated for
E3