32 Revision 02, January 2011
8. M
AINTENANCE
8.1. Cleaning
Instructions
Caution:
Before cleaning the handpiece and main unit, always turn the unit off,
disconnect it from energy and wait a minimum of 5 minutes.
Do not use cleaning solutions that contain ingredients that may harm
plastic, such as ammonia, acetone, hydrochloric acid (HCI), chlorine
bleach etc.
8.1.1.
Lamp & Reflector
During treatment, debris accumulates on the lamp. This may affect
performance and the life of the lamp. Use the supplied cleaning tool to clean
the lamp after each treatment (Figure 19).
Figure 19: Handpiece Cleaning Tool
1. Put 2 drops of alcohol on the sponge part of the cleaning tool (Figure 20)
Figure 20: Applying Alcohol to the Cleaning Tool
2. Use Tip B to gently rub along the length of the lamps
between the two
lamps
until the debris is removed (Figure 21).
Tip B
Use this end to clean
between the lamps
Tip A
Use this end to clean between
the lamps and reflector.