3152A User Manual
3152A Specifications A-14
Racal Instruments
©
2001
Sequenced
Waveforms
Operation
Permits division of the memory bank
into
smaller
segments.
Segments
may be linked and repeated in user-
selectable
fashion
to
generate
extremely
long
waveforms
Sequencer steps
From 0 to 4096
Segment loops
From 0 to 1 million.
Segment Duration
Minimum 100 ns for more than one
loop
Sampling Clock
Source
Internal
synthesizer,
external
clock,
ECLTRG0
Range
Internal
100 mHz to 100 MHz
External To
100
MHz
Internal Reference
10 MHz, fixed internal crystal clock
frequency
Resolution
7
digits
Accuracy
Same as frequency standard
Stability
Same as frequency standard
Swept Waveforms
Operation
Utilizes the 3152A’s sequencer and
waveform memory to implement a
frequency
sweep
from
one
frequency
to another. Note: Sweep mode is
volatile
in
that
any
arbitrary
waveforms
or
sequences
stored
in
memory are deleted when sweep is
invoked.