The synthesizer returns a string that contains all the parameters
associated with the square function similar to the following string:
“1.6,2.000,0.000,50"
The command:
APPLy:
SINusoid
{<frequency>,<amplitude>,<offset>,
<phase>,<power>}
programs the synthesizer to output a sine waveform with
frequency, amplitude, offset, start phase and power parameters.
Parameters are not optional if the above APPLy command is
used. Include all other parameters in the command. The default
settings for these functions are: 1 MHz, 5 Vp-p, 0 V, 0 and 1.
APPLy: TRIangle {<frequency>,<amplitude>,<offset>,
<phase>,<power>}
programs the synthesizer to output a triangle waveform with
frequency, amplitude, offset, start phase, and power parameters.
The default settings for these functions are: 1 MHz, 5 Vp-p, 0 V,
0 and 1.
The
command:
APPLy:
SQUare
{<frequency>,<amplitude>,<offset>,
<duty_cycle>}
programs the synthesizer to output a square waveform with
frequency, amplitude, offset and duty cycle parameters. The
default settings for these functions are: 1 MHz, 5 Vp-p, 0 V, and
50%.
The command:
APPLy:
PULSe{<frequency>,<amplitude>,<offset>,
<delay>,<high_time>,<rise_time>,<fall_time>}
programs the synthesizer to output a pulse waveform with
frequency, amplitude, offset, delay, rise time, high time and fall
time parameters. The default settings for these functions are: 1
MHz, 5 Vp-p, 0 V, 0%, 10%, 10% and 10%.