3152A User Manual
Maintenance And Performance Checks 5-6 Racal Instruments
©
2001
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP
SQU
:FREQ 10
:VOLT
1
2.
Set the counter to frequency measurement.
3.
Connect the Model 3152A output to the counter
=
s input.
Change the frequency setting as required for the test
and verify the reading on the counter display as follows:
3152A SETTING
COUNTER READING
10.00000 Hz
9.999000 Hz - 10.00100 Hz
100.0000 Hz
99.99000 Hz - 100.0100 Hz
1.000000 KHz
999.9000 Hz - 1.000100 KHz
10.00000 KHz
9.999000 KHz - 10.00100 KHz
100.0000 KHz
99.99000 KHz - 100.0100 KHz
1.000000 KHz
999.9000 KHz - 1.000100 KHz
10.00000 MHz
9.999000 MHz - 10.00100 MHz
50.00000 MHz
49.99500 MHz - 50.00500 MHz
Amplitude
Accuracy
Accuracy Specifications (1 KHz):
±
(1% of r200
μ
V), to 160 mV
±
(1% of r2 mV), to 1.6 V
±
(1% of r20 mV), to 16 V
Equipment
:
DMM, 50
Ω
feedthrough termination
1.
Configure the Model 3152A as follows:
Function:
Sine
Frequency:
1.000 KHz
Amplitude:
15 V