3152A User Manual
Racal Instruments
©
2001
Maintenance and Performance Checks 5-9
Set the oscilloscope and verify that the rise and fall times are
less than 6 ns. Verify that overshoot and ringing are less than
5% of amplitude.
Sine
Characteristics
Specified Total Harmonic Distortion:
<0.1% from 10.00Hz to
100.0KHz (with a 4000 point waveform).
Specified Harmonic Signals:
<5MHz,
<10Vp-p, -50dBc
<5MHz,
<16Vp-p, -45dBc
<10MHz,
<10Vp-p, -40dBc
<10MHz,
<16Vp-p, -35dBc
<50MHz,
<10Vp-p, -28dBc
<50MHz,
<16Vp-p, -23dBc
Equipment:
Krohn-Hite 6900 (or equivalent) Distortion
Analyzer, RF Spectrum Analyzer, 50
Ω
feedthrough termination
(or 20dB 50
Ω
attenuator if required).
1.
Connect the Model 3151 output to distortion analyzer
input.
2.
Set the distortion analyzer to distortion measurements,
set the Model 3151 frequency setting, and verify the
distortion reading as below.
3.
Configure the Model 3151 as follows:
Function:
USER Waveform
Frequency:
10.00Hz
Amplitude:
5V
Remote Commands ::RES
:OUTP ON
:FUNC:MODE
USER
:TRAC:DEF
4000
:TRAC:SEL 1
:FREQ:RAST
40e3
:VOLT5