3152A User Manual
Maintenance And Performance Checks 5-12 Racal Instruments
©
2001
external pulse generator connected to the Model 3152A
for the next test.
Gate
1.
Configure the Model 3152A as follows:
Function:
Sine wave
Frequency:
1.000 KHz
Operating Mode:
Gated
Remote Commands: :RES
:OUTP ON
:FREQ
1e3
:INIT:CONT
OFF
:GATE:STAT
ON
Set the oscilloscope and verify that the Model 3152A outputs a
gated signal. Leave the external pulse generator connected to
the Model 3152A for the next test.
Burst
1.
Configure the Model 3152A as follows:
Function:
Sine wave
Frequency:
1.000 KHz
Operating Mode:
Burst
Remote Commands: :RES
:OUTP ON
:FREQ
10e3
:INIT:CONT
OFF
:BURST:STAT ON
:TRIG:COUN 10
2. Set the oscilloscope and verify that the Model 3152A
outputs a burst of 10 complete output waveforms.
Remove the external pulse generator connection from
the Model 3152A for the next test.