![Racal Instruments 1256E Manual Download Page 156](http://html1.mh-extra.com/html/racal-instruments/1256e/1256e_manual_3132597156.webp)
1256E User Manual
Digital Module Operation 7-10
Racal Instruments
©
2002
After executing the command:
DIG:SYNC:DATA
(@7(4)),50,60,70,80,90
the memory buffer for port 4 of module 7 tracks that depicted in
Figure 7-1 (B). There are now a total of nine bytes loaded. The
index for this port now points to the next available position in the
buffer.
After executing the command:
DIG:SYNC:INDEX
(@7(4)),3
the index for the port buffer now points to the fourth entry in the
buffer. This is where the next data will be loaded. This is depicted
in Figure 7-1 (C).
After executing the command:
DIG:SYNC:DATA
(@7(4)),77,78
the fourth and fifth data bytes are overwritten. The previous data
(40 and 50) is replaced by the new data (77 and 78). The index
now points to the sixth entry, which has a value of 60. The next
DIG:SYNC:DATA command would begin by overwriting this data.
This final configuration is depicted in Figure 7-1 (D).