Page
55
of
133
Version 3.0 REV r000
Date 25-08-2010
6.5.3
Time Variable Gain
Absorption and spreading loss are the main components of the Time Variable Gain (TVG)
computation.
TVG Equation
TVG = 2*R*a/1000 + Sp*log(R) + G
α
= Absorption Loss db/km
R = Range in metres
Sp = Spreading loss coefficient
G = Gain from Sonar Control setting
TVG is employed in underwater acoustics to compensate for the nature of the reflected acoustic
energy. When an acoustic pulse is transmitted in a wide pattern, the first returns will generally be
from the nadir region and very strong. As the receive window time lengthens, the weaker returns
are received. Using a fixed gain would apply either too much gain for the early returns or
insufficient gain for the later returns. The solution is to use TVG. The function of TVG is to increase
gain continuously throughout the receive cycle. Therefore, smaller gain corresponds with the first
returns (normally the strongest) and higher gain corresponds to the later returns (normally the
weakest). This function is represented in, what is called, the TVG curve.
6.5.3.1
TVG Curve
The TVG curve can be either shallow or steep depending mostly on the Absorption value to define
the shape of the curve. The Spreading Loss will determine the amplitude of the gain.
Figure 50: TVG Curve Concept
Summary of Contents for 2022
Page 1: ...SONIC 2024 2022 BROADBAND MULTIBEAM ECHOSOUNDERS Operation Manual V3 0 Part No 96000001 ...
Page 2: ...Page 2 of 133 Version 3 0 Rev r000 Date 25 08 2010 Part No 96000001 ...
Page 92: ...Figure 78 Smooth log information copied from real time survey log ...
Page 96: ...Page 96 of 133 Version 3 0 Rev r000 Date 25 08 2010 This page intentionally left blank ...
Page 112: ...Page 112 of 133 Version 3 0 Rev r000 Date 25 08 2010 This page intentionally left blank ...
Page 116: ...Page 116 of 133 Version 3 0 Rev r000 Date 25 08 2010 This page intentionally left blank ...
Page 125: ...Page 125 of 133 Version 3 0 REV r000 Date 25 08 2010 Figure 94 Sonic 2024 Receive Module ...