MedTrack CR60 Rehabilitation Treadmill Service Manual
Preventative Maintenance 5-3
Caution! Use extreme caution to avoid physical damage to the
internal components. Avoid static discharge, which can
damage the electronic components.
1. Remove the hood as described on “Removing the Treadmill Hood”
2. Vacuum the areas behind the AC Drive Module. Do not vacuum the
drive board.
3. Replace the hood.
Replacement Schedules
Replace belts if they are frayed or show excessive wear.
Electrical Testing
Electrical testing is to be done by the facility’s biomedical department as
required. Check the leakage current of the treadmill periodically—at least
every nine months—to be sure it does not exceed local standards. See
Appendix B, Specifications, for limitations.