Motor Armature –
approximately 5
ohms.
Switch pressed – Open
Switch not pressed – Closed
Switch pressed – Closed
Switch not pressed – Open
Verify correct operation of the 18-degree reset
3.
switch by checking its operation at the 6-pin
connector. Refer to the following drawing for the
correct readings.
Replace the 18-degree reset switch assembly if
4.
defective and retest. Re-home the seating system
by fully operating the tilt or using the HHP and
retest.
Note: Contact Sunrise Medical Tech Support for as-
sistance.