SEATING SECTION
Verify that the BUS cables leading to the seating
3.
system are mated at the back of the wheelchair.
If disconnected, reconnect and retest.
Check for battery voltage at the 2-pin connector
4.
leading to the QR-MAC
If battery voltage is present at the 2-pin connector
5.
and no seating functions are illuminated, replace
the QR-MAC and retest.
If battery voltage is not present, locate the 2-pin
6.
connector leading from the base to the seating
system
Fig 3.64
Fig 3.65
Fig 3.66