rst step in Troubleshooting any Power Wheelchair is to look at the chair and break it down into four simpler
groups.
Refer to the following tables for a quick diagnostic
guide provided by the Battery Gauge on the Quickie
system.For additional information, refer to the Users
guide for each system.
Priority Values show which error takes priority in the
event of multiple errors.
(1=Highest Priority, 10=Lowest Priority)
Drive Mode errors are indicated by the red LED next to
the Mode Button Flashing RED.
Module
Motors
Harness
Battery
LED illumination state
Error Description
Priority *
Motor Controller Internal
Module Error
2
Module Communication Error
3
Input Device Out of Neutral at
Power On
8
Park Brake Open Circuit Error 6
Right Motor Open Circuit Error
or Right Motor Encoder Error
4
Left Motor Open Circuit Error or
Left Motor Encoder Error
5
Battery Under Voltage Error or
Battery Over Voltage Error
9
Motor Controller High Tempera-
ture Warning
10
Invalid System Con
fi
guration
Error
1
Drive Lockout External Source 7
Drive Mode Errors
Diagnostics Overview
Quickie by Delphi Diagnostic Codes
Summary of Contents for Rhythm
Page 1: ......
Page 2: ...2 2006 Sunrise Medical...
Page 14: ...14 2006 Sunrise Medical INTRODUCTION...
Page 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL...
Page 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2...
Page 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL...
Page 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL...
Page 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL...
Page 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8...
Page 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 181: ......
Page 182: ...182 2006 Sunrise Medical...