23
©2006 Sunrise Medical
POWER BASE SECTION
Motors
Brake Solenoid
Motor
Motor Connector
Fig 1.19
Fig 1.20
Motor
Motor
Resistance
Brake Solenoid
Resistance
Standard
0-2 ohms
21.0-22.0 ohms
Encoder
0-2 ohms
15.0-17.0 ohms
Checking the Motor Brushes
Brush Assembly
Wire between Coil
Top Solder
Commutator Inspection
Fig 1.21
Fig 1.22
Fig 1.23
Fig 1.24
Note: When removing brushes from the motor for
inspection, please note orientation and location of the
brushes as they are removed from the motor (
fi
g 1.21).
The brushes are “burned in” to the commutator and
reinstallation in a location or orientation not matching
the pre-inspection location may negatively affect motor
operation.
Checks:
• How smooth is the brush surface – did it create “C”
shaped groove?
• If there is less than 1/4” brush material left the
Brushes should be replaced.
• How the wire between the coil looks – did it discolor
(
fi
g 1.22)
• Did the Top soldered joint melt (
fi
g 1.23).
• How does the commutator look after the brush has
been contacting the surface (
fi
g 1.24).
Checking Motor Resistance and Continuity
Fig 1.18
Summary of Contents for Rhythm
Page 1: ......
Page 2: ...2 2006 Sunrise Medical...
Page 14: ...14 2006 Sunrise Medical INTRODUCTION...
Page 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL...
Page 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2...
Page 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL...
Page 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL...
Page 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL...
Page 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8...
Page 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 181: ......
Page 182: ...182 2006 Sunrise Medical...