35
©2006 Sunrise Medical
POWER BASE SECTION
Adjusting the timing of the anti-pitch effect is
•
accomplished by rotating the nut at the end of
the anti-pitch actuation cable at the anti-pitch
housing. The anti-pitch housing is located on
the stainless steel anti-pitch shaft inboard of
the drive motor. (Fig 1.63) (NOTE: The free
threads seen between the nut and the cable
rotary barrel indicate that this anti-pitch is fully
engaged.)
As seen, the barrel end of the cable has
•
a hexagonal shape and while an adjustable
wrench could be set to
fi
t this portion of the
cable, a vise-grip type of pliers would be better
suited to this application.
Adjusting the nut clockwise will increase the
•
amount of suspension travel prior to anti-pitch
damping engagement. Adjusting the nut counter
clockwise will reduce the amount of suspension
travel prior to anti-pitch engagement.
After evenly adjusting the timing of the anti-pitch
•
on each side, reinstall the 6mm anti-sway bar
and attaching hardware, then set torque to 6
ft-lbs or 8.13 Newton-meter.
Remove the chair from the pedestal. The
•
procedure is complete.
Cable Attach & Adjust
Fig 1.63
Summary of Contents for Rhythm
Page 1: ......
Page 2: ...2 2006 Sunrise Medical...
Page 14: ...14 2006 Sunrise Medical INTRODUCTION...
Page 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL...
Page 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2...
Page 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL...
Page 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL...
Page 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL...
Page 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8...
Page 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 181: ......
Page 182: ...182 2006 Sunrise Medical...