68
©2006 Sunrise Medical
QUICKIE ELECTRONICS SECTION
Actuator Test (QMAC controlled actuators)
Complete the following steps to test the seating system
actuators.
Connect the Quickie iQ or HHP to the wheel-
1.
chair.
Turn on the wheelchair. Prior to completing any
2.
test, re-home the wheelchair using either the HHP
or the PCSS.
If using the HHP, select Monitor, then QMAC Ac-
3.
tuators. Under QMAC Actuators, select the actua-
tor to be tested (i.e., Tilt). (
fi
g2.23.26)
If using the Quickie iQ, select Monitor Parameter,
4.
followed by selecting QMAC, then actuator to test
(i.e., Tilt). (
fi
g 2.23.27)
Note the position of the actuator (counts and
5.
degrees). Operate the actuator and observe the
counts, degrees and current. Refer to Table 1 for
a listing of typical currents for each actuator.
Tilt
+ + -
Angle
3.0 deg
Counts
0
Current
0.000 A
SW2 Status
open
Actuator
Average Current – No Load
Maximum Current –Under Load
Recline
1 to 2 amps
8 amps
Tilt
1 to 2 amps
8 amps
Lift
1 to 2 amps
8 amps
Left Leg
1 to 2 amps
4 amps
Right Leg
1 to 2 amps
4 amps
Power Shear
1 to 2 amps
4 amps
Fig 2.23.26
Fig 2.23.27
Table 1
Summary of Contents for Rhythm
Page 1: ......
Page 2: ...2 2006 Sunrise Medical...
Page 14: ...14 2006 Sunrise Medical INTRODUCTION...
Page 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL...
Page 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2...
Page 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL...
Page 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL...
Page 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL...
Page 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8...
Page 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Page 181: ......
Page 182: ...182 2006 Sunrise Medical...